Ces radiol. 2013, 67(3):232-237

Exposure parameters of cine mode and fluoroscopy mode of the angiography systemOriginal article

Lucie Súkupová
Institut klinické a experimetnální medicíny, Praha

Aim: The aim of this work was to analyse exposure parameters of cine mode and fluoroscopy mode and therefore to understand the automatic exposure control of an angiography system used for interventional cardiology procedures. The influence of the additional filtration on dose to patients was assessed.

Methods: The study was performed on the angiography system Axiom Artis MP with an image intensifier. Exposure parameters selected by the automatic exposure control of many cine modes for adult patients as well as for pediatric patients were analysed. Mainly it included tube voltage, tube current, pulse width and additional filtration.

Results: The main difference between cine mode and fluoroscopy mode was in used tube current and pulse width, smaller difference was between additional filtration. With increasing thickness of the phantom, the AEC makes the tube current increase, pulse width and voltage is kept constant. With still increasing phantom width, current and pulse width increase, voltage is kept constant. Then tube current, pulse width and also voltage increase. With voltage increased, the additional filtration is lowered.

Conclusion: The exposure parameters of cine mode were a little bit different from parameters of fluoroscopy mode. Thicker additional filtration significantly lowers the skin dose to patients but cannot be used for obese patients due to excessive X-ray tube loading.

Keywords: angiography system, additional filtration, exposure parameters, skin dose

Accepted: March 15, 2013; Published: September 1, 2013  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Súkupová L. Exposure parameters of cine mode and fluoroscopy mode of the angiography system. Ces radiol. 2013;67(3):232-237.
Download citation

References

  1. American Association of Physicists in Medicine. Functionality and operation of fluoroscopic automatic brightness control/automatic dose rate control logic in modern cardiovascular and interventional angiography systems. A report of AAPM Task group 125, 2012.
  2. International Atomic Energy Agency. Radiation protection in paediatric interventional cardiology. Lecture 10. Výukové materiály IAEA pro intervenční kardiologii.
  3. Nicholson R, Tuffee F, Uthappa MC. Skin sparing in interventional radiology: The effect of copper filtration. Br J Radiol 2000; 73: 36-42. Go to original source... Go to PubMed...
  4. Súkupová L. Radiační zátěž pacientů v intervenční kardiologii. Dizertační práce. Praha: ČVUT, Fakulta jaderná a fyzikálně inženýrská 2012.

This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), which permits use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.